Qingdao Junray Intelligent Instrument Co., Ltd.

produkten

Masker Bakteriële Filtration Effektiviteit (BFE) Tester

Koarte beskriuwing:

Key specifications of Model ZR-1000 mask bacteria filtration efficiency tester (BFE) not only meet the requirements of B.1.1 in YY0469-2011 <technical requirements for medical surgical masks>, but also meet the requirements of ASTMF2100, ASTMF2101 and European EN14683 standards.


Produktdetail

FAQ

Produkt tags

priis:US $ 48,000-90,000 / stik

Ynlieding

Key specifications of Model ZR-1000 mask bacteria filtration efficiency tester (BFE) not only meet the requirements of B.1.1 in YY0469-2011 <technical requirements for medical surgical masks>, but also meet the requirements of ASTMF2100, ASTMF2101 and European EN14683 standards. What’s more, we made innovative improvements based on this, with double gas contrast sampling method at the same time, improve the precision of sampling. It is applicable to the performance test of the mask bacterial filtration efficiency by the metrological verification department, scientific research institutes, mask manufacturers and other relevant departments.

Funksjes

It testsysteem foar negative druk soarget foar de feiligens fan operators;

De kast mei negative druk is foarsjoen fan peristaltyske pomp, A en B twa-paad seis poadium Anderson sampler;

De stream fan peristaltyske pomp kin ynsteld wurde;

De streamen fan mikrobiële aerosolgenerator kinne wurde ynsteld.

Goede ferneatigingsprestaasjes;

Kontroleare troch ynbêde industriële mikrocomputer mei hege snelheid;

10.4-inch yndustriële hege helderheidskleurich oanraakskerm;

USB-poarte, stipe gegevensferfier fan USB-flashdisk;

Ynboude ljochten mei hege helderheid;

Ynboude skeakel foar lekkage-beskerming om de feiligens fan operators te beskermjen;

De binnenste laach fan 'e kast is makke fan rustfrij stiel, en de bûtenste laach is kâldgewalst blêd spuite troch plestik;

Waarmbehâld en flamfertrager wurdt ynsteld tusken binnen- en bûtenlagen;

De foar iepen te iepenjen glêzen doar is handich foar de eksperimintele operator om te observearjen en te operearjen;

Afneembare beugel mei ferstelbere hichte;

De caster kin brûkt wurde foar stipe en ferpleatse.

Noarmen

EN14683-2019

ASTM F2100, STM F2101

YY 0469-2011

YY / T 0969-2013

GB / T 38880-2020

T / CNTAC 55—2020 T / CNITA 09104——2020

T / CTCA 7—2019

Spesifikaasjes

Wichtichste parameters

 Parameter berik 

Resolúsje

MPE (maksimum tastiene flater)

Sampling Flow Rate fan A. 28.3L / min 0.1L / min ± 2,5%
Sampling Flow Rate of B 28.3L / min 0.1L / min ± 2,5%
Nebulisaasjestream (8 ~ 10) L / min 0.1L / min ± 5,0%
Peristaltyske pompestream

(0,006 ~ 3,0) ml / min

0.001mL / min ± 2,5%
Foarste druk fan A-paad flowmeter (-20 ~ 0) kPa 0.1kPa ± 2,5%
Foardruk fan B-paad flowmeter (-20 ~ 0) kPa 0.1kPa ± 2,5%
Foardruk fan ferstuiver (0 ~ 300) kPa 0.1kPa ± 2,5%
Negative druk yn aerosolkeamer (-90 ~ -120) Pa 1Pa ± 2,0%
Operaasjetemperatuer (0 ~ 50) ℃
Kabinet negative druk (-50 ~ -200) Pa
Fermogen foar gegevensopslach > 100000 groepen
Skaaimerken fan HEPS filter Filtraasje-effisjinsje foar dieltsjes is boppe 0.3μm ≥ 99.99%
Gemiddelde massadiameter fan aerosols Gemiddelde partikeldiameter is (3,0 ± 0,3) μm, geometryske standertdeviaasje ≤ 1,5
Fongen dieltsje grutte fan twa-paad 6-poadium Anderson sampler 1e etappe> 7 μm, 2e etappe (4.7-7) μm, 3e etappe (3.3-4.7) μm, 4e etappe (2.1-3.3) μm, 5e etappe (1.1-2.1) μm, 6e etappe (0.6-1.1) μm
Grutte fan spuitkeamer (lingte 600 × diameter 85 × dikte 3) mm
Totaal oantal positive sampler foar kwaliteitskontrôle (2200 ± 500) CFU
Ventilaasjestream fan kast mei negative druk ≥5m3/ min
Ofmjittingen fan kastdoar mei negative druk (lingte 1000 × breedte 730) mm
Hostgrutte (lingte 1180 × breedte 650 × hichte 1300) mm
Beugelgrutte (lingte 1180 × breedte 650 × hichte 600) mm, hichte ferstelber binnen 100mm
Streamtafier AC220V ± 10%, 50Hz
Ynstrumint lûd < 65dB (A)
Gewicht fan heule masine Likernôch 250 kg
Algemien enerzjyferbrûk < 1500W

  • Foarige:
  • Folgjende:

  • Skriuw jo berjocht hjir en stjoer it nei ús